Recalls

Abbott Diabetes Care: Class 1 Recall – FreeStyle InsuLinx Blood Glucose Meters – Risk of Incorrect Test Result

Silver Spring, Maryland–(ENEWSPF)–May 20, 2013.

Proud member of LION Publishers

ISSUE: Abbott initiated a voluntary recall of FreeStyle lnsulinx Blood Glucose Meters in the United States. At extremely high blood glucose levels of 1024 mg/dL and above, the FreeStyle InsuLinx Blood Glucose Meter will display and store in memory an incorrect test result that is 1024 mg/dL below the measured result. Blood glucose levels at 1024 mg/dL and above are very rare. However, if high blood glucose levels of 1024…

Most read stories this week

Take a Survey

ARCHIVES