Recalls

Abbott Initiates Voluntary Recall of FreeStyle lnsulinx® Blood Glucose Meters in the United States

At rare, extremely high blood glucose levels (1024 mg/dL and above), the FreeStyle lnsuLinx Blood Glucose Meter may provide an inaccurate reading

Silver Spring, Maryland–(ENEWSPF)–April 17, 2013.  Abbott has announced it is initiating a voluntary recall of FreeStyle lnsulinx® Blood Glucose Meters in the United States.

Proud member of LION Publishers

The company has determined that at extremely high blood glucose levels of 1024 mg/dL and above, the FreeStyle lnsulinx Meter will display and store in memory an incorrect test…

Most read stories this week

Take a Survey

ARCHIVES